smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter — Class II medical device recall Z-0466-2025

Ongoing recall by Smiths Medical ASD, Inc., reported 2024-11-27. The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the b

Recall numberZ-0466-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSmiths Medical ASD, Inc., Minneapolis, MN, United States
Recall initiated2024-06-26
Classified2024-11-15
Reported by FDA2024-11-27
Quantity156,400 ac
Model numbers21-0270-25

Product

smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps

Reason for recall

The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.