Data Foundry

Medical device recalls 2012

Stryker Orthopaedics FEMORAL OFFSET BUSHING TRITHION REVISION… — Class II medical device recall Z-0469-2013

Terminated recall by Stryker Howmedica Osteonics Corp., reported 2012-12-12. Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Ti

Recall numberZ-0469-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Howmedica Osteonics Corp., Mahwah, NJ, United States
Recall initiated2012-07-27
Classified2012-12-04
Reported by FDA2012-12-12
Terminated2015-09-30
Quantity7 units
Lot numbersNE1WR1
Model numbers6543-1-600

Product

Stryker Orthopaedics FEMORAL OFFSET BUSHING TRITHION REVISION INSTRUMENTS REF 6543-1-600;NON-STERILE Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA A Subsidiary of Stryker Corp. Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 Meyzieu Cedex France The Offset Bushings are parts of the Triathlon TS system. They are used during trialing to assess the appropriate magnitude and orientation of offset between the Implant and the intramedullary Stem. The feedback from the Offset Bushing allows for the selection of the corresponding Offset Adaptor, which allows for optimal bone coverage, compensating for differences between the Implant and bone canal axes. Made in USA

Reason for recall

Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset Bushing did not include rotational markings.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0469-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.