NexGen¿ Complete Knee Solutions Stemmed Tibial Component Precoat… — Class II medical device recall Z-0480-2014
Terminated recall by Zimmer, Inc., reported 2013-12-18, distributed nationwide. Two complaints were received in which it was reported that the stem would not lock into the tibia. Evaluation
| Recall number | Z-0480-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zimmer, Inc., Warsaw, IN, United States |
| Recall initiated | 2013-10-31 |
| Classified | 2013-12-11 |
| Reported by FDA | 2013-12-18 |
| Terminated | 2015-04-30 |
| Distributed | Nationwide (US) |
| Countries outside the US | CN, JP, TW |
| Quantity | 128 units |
| Lot numbers | 6228417 |
Product
NexGen¿ Complete Knee Solutions Stemmed Tibial Component Precoat NexGen Tibial Baseplates are part of the NexGen systems of semi-constrained, non-linked condylar knee prosthesis. The device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis; Collagen disorders, and/or avascular necrosis of the femoral condyle; Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; Moderate valgus, varus, or flexion deformities.
Reason for recall
Two complaints were received in which it was reported that the stem would not lock into the tibia. Evaluation of the returned device found that the stemmed portion of the tibial plate would not hold the stem as the connection was too loose. The taper diameter was found to be oversized by .010.