The Artis zee / zeego Angiography System is designed as a set of… — Class II medical device recall Z-0486-2015

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2014-12-10, distributed nationwide. There is a potential issue on running Artis systems running software VC1x software if a network problem arises

Recall numberZ-0486-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2013-09-11
Classified2014-11-28
Reported by FDA2014-12-10
Terminated2015-03-30
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, CH, CL, CN, CO, CZ, DE, DK, EG, ES, FI, FR, GB, HK, HU, ID, IE, IL, IN, IT, JO, JP, KE, KR, MY, NL, NO, NZ, PH, PK, PL, PT, RS, SA, SE, SG, SK, TH, TR, TW, UA, VN, ZA
Quantity1,131
Reason classessoftware
Serial numbers135000, 135102, 135109, 135111, 135112, 135114, 135117, 135120, 135122, 135124, 135125, 135129, 135132, 135133, 135137, 135143, 135144, 135145, 135148, 135151, 135152, 135157, 135169, 135179, 135180 and 475 more
Model numbers10094135, 10094137, 10094139, 10094141, 10094142, 10094143, 10280959

Product

The Artis zee / zeego Angiography System is designed as a set of components that may be combined into different configurations to provide specialized angiography systems.

Reason for recall

There is a potential issue on running Artis systems running software VC1x software if a network problem arises, the function cannot be deactivated again by pressing the "Block Radiation" key on the touch screen control. If this behavior occurs, the system image generation function is not available for patient examinations without switching the system off and back on again manually. This