Data Foundry

Medical device recalls 2014

McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50… — Class II medical device recall Z-0494-2015

Terminated recall by Aircraft Medical Limited, reported 2014-12-10, distributed nationwide. The blade retaining clip may be difficult to engage and thereby increase the possibility of a blade becoming d

Recall numberZ-0494-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAircraft Medical Limited, EDINBURGH, N/A, United Kingdom
Recall initiated2014-11-04
Classified2014-12-02
Reported by FDA2014-12-10
Terminated2017-11-17
DistributedNationwide (US)
Countries outside the USAT, CN, DK, FI, FR, HK, IL, IT, KR, NL, PL, SA, SG, TR, TW
Lot numbers13042401, 13121202, 13122301, 14030501, 14031814, 14061616, 2016/12/11, 2016/12/25, 2017/3/16, 2017/3/26, 2017/5/5, 2017/6/25

Product

McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50 items, each individually peel-pouched and EtO sterilized. Product Usage: Single Use, sterile video laryngoscope blade for use with McGRATH MAC Video Laryngoscopes

Reason for recall

The blade retaining clip may be difficult to engage and thereby increase the possibility of a blade becoming disconnected during intubation, specifically when the device is being retracted from the patients mouth.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0494-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.