KIT URIN CUP 16X100 8 — Class II medical device recall Z-0494-2019
Terminated recall by Becton Dickinson & Company, reported 2018-11-28, distributed in IL. BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when u
| Recall number | Z-0494-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Becton Dickinson & Company, Franklin Lakes, NJ, United States |
| Recall initiated | 2018-08-30 |
| Classified | 2018-11-21 |
| Reported by FDA | 2018-11-28 |
| Terminated | 2020-10-21 |
| Distributed | IL |
| Model numbers | 364989BD |
Product
KIT URIN CUP 16X100 8.0 CONI UA YEL, Cat. No. 364989 BD Vacutainer Urine Cup Kit with UA Tube for Midstream Specimens
Reason for recall
BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0494-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.