Data Foundry

Medical device recalls 2015

DrugSmart 12 Test Cup — Class III medical device recall Z-0495-2016

Terminated recall by Ameditech Inc, reported 2015-12-30, distributed nationwide. Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (

Recall numberZ-0495-2016
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAmeditech Inc, San Diego, CA, United States
Recall initiated2015-11-16
Classified2015-12-24
Reported by FDA2015-12-30
Terminated2016-11-09
DistributedNationwide (US)
Quantity160 kits
Model numbers61205D, 61210D

Product

DrugSmart 12 Test Cup (Multi & BUP), Item No. 61210D DrugSmart 12 Test Cup (Multi/Single/BUP), Item No. 61205D

Reason for recall

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0495-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.