Medtronic SynchroMed II — Class I medical device recall Z-0497-2013
Terminated recall by Medtronic Neuromodulation, reported 2012-12-19, distributed nationwide. Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the Synchr
| Recall number | Z-0497-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Neuromodulation, Minneapolis, MN, United States |
| Recall initiated | 2012-11-09 |
| Classified | 2012-12-13 |
| Reported by FDA | 2012-12-19 |
| Terminated | 2017-12-31 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, MX |
| Quantity | 174,700 devices |
| Reason classes | unapproved product |
| Model numbers | 8637 |
Product
Medtronic SynchroMed II, Model 8637, (The SynchroMed¿ II Pump is supplied in 20 ml or 40 ml reservoir size.) Sterilized using ethylene oxide. The pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implantable components of the SynchroMed II Infusion System include the pump, catheter, and catheter accessories.
Reason for recall
Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps.