Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with… — Class II medical device recall Z-0499-2026
Ongoing recall by SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION, reported 2025-11-19, distributed nationwide. Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could pot
| Recall number | Z-0499-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION, Kyoto, Japan |
| Recall initiated | 2025-10-16 |
| Classified | 2025-11-13 |
| Reported by FDA | 2025-11-19 |
| Distributed | Nationwide (US) |
| Countries outside the US | HT |
| Quantity | 1,766 devices |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 04540217014484, 04540217050826 |
| Serial numbers | 0162M42920, 0162M43015, 0162M43016, 0162M43017, 0162M43218, 0162M43518, 0162M43610, 0162M43611, 0162M43612, 0162M43613, 0162M43614, 0162M43615, 0162M43616, 0162M43617, 0162M43618, 0162M43619, 0162M43620, 0162M43916, 0162M44409, 0162M44509, 0162M44510, 0162M45101, 0162M45102, 0162M45118, 0162M45119 and 475 more |
| Model numbers | CH-200, CH-200M |
Product
Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the X-ray tube and collimator to position the X-ray tube unit. This unit is used for radiographic examinations of whole body except mammography when combined with an X-ray high voltage generator, an X-ray tube unit, an X-ray collimator, and if necessary, an X-ray radiography stand, an X-ray radiography table, and digital radiography system.
Reason for recall
Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating shaft.