MOSAIQ Oncology Information System Software Builds 3 — Class II medical device recall Z-0503-2026
Ongoing recall by ELEKTA SOLUTIONS AB, reported 2025-11-26, distributed nationwide. When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of t
| Recall number | Z-0503-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | ELEKTA SOLUTIONS AB, Stockholm, Sweden |
| Recall initiated | 2025-09-17 |
| Classified | 2025-11-14 |
| Reported by FDA | 2025-11-26 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, MT, ZA |
| Quantity | 4 systems |
| UPC / GTIN / UDI-DI | 07340201500026, 07340201500071 |
Product
MOSAIQ Oncology Information System Software Builds 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 Software Version and UDI Codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500026(10)3.1.3.2 3.1.3.3 (01)07340201500026(10)3.1.3.3 3.1.3.4 (01)07340201500026(10)3.1.3.4 3.2.1.0 (01)07340201500071(10)3.2.1.0 3.2.1.1 (01)07340201500071(10)3.2.1.1 3.2.1.2 (01)07340201500071(10)3.2.1.2 3.2.1.3 (01)07340201500071(10)3.2.1.3 3.2.1.4 (01)07340201500071(10)3.2.1.4 3.2.2.0 (01)07340201500071(10)3.2.2.0 3.2.2.1 (01)07340201500071(10)3.2.2.1 3.2.3.0 (01)07340201500071(10)3.2.3.0 MOSAIQ¿ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.
Reason for recall
When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule.