Data Foundry

Medical device recalls 2012

WANG Transbronchial Aspiration Needles Cat — Class II medical device recall Z-0505-2013

Terminated recall by ConMed Corporation, reported 2012-12-19, distributed in CA, FL, IA, IN, MA, MD, ME, MO…. ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration Needles woul

Recall numberZ-0505-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmConMed Corporation, Utica, NY, United States
Recall initiated2012-11-20
Classified2012-12-10
Reported by FDA2012-12-19
Terminated2015-08-25
DistributedCA, FL, IA, IN, MA, MD, ME, MO, MS, NC, NM, OH, PA, TX
Countries outside the USES, HU, JP, PT
Quantity452
Lot numbers1209171, 1209201
Model numbersMW-221

Product

WANG Transbronchial Aspiration Needles Cat. No. MW-221, WANG Cytology Needles, Central or Peripheral Regions, 21 g x 15 mm long needle. --- Common Name: Bronchoscope (flexible or rigid) and accessories. The ConMed WANG Transbronchial Aspiration Needles are designed for use with a bronchoscope. They are used to puncture the tracheobronchial wall and to aspirate sufficient tissue and/or cell specimens to stage bronchogenic carcinoma. These are disposable sterile, single-use devices.

Reason for recall

ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration Needles would not insert properly through the instrument channel of a 2.0 mm bronchoscope. If the device will not fit through the instrument channel of a 2.0 mm bronchoscope, it must be replaced with an alternate device and could result in minor inconvenience to the user.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0505-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.