Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular… — Class II medical device recall Z-0505-2026
Ongoing recall by Thoratec LLC, reported 2025-11-26, distributed nationwide. Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup B
| Recall number | Z-0505-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Thoratec LLC, Pleasanton, CA, United States |
| Recall initiated | 2025-10-09 |
| Classified | 2025-11-14 |
| Reported by FDA | 2025-11-26 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BE, BG, BH, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GR, IE, IL, IN, IQ, IR, IT, JP, KR, KW, KY, LB, LT, MA, MY, NL, NO, NZ, OM, PE, PL, PT, QA, RO, RS, RU, SA, SE, SG, SI, TH, TN, TR, TW, ZA |
| Quantity | 29,059 |
| UPC / GTIN / UDI-DI | 00813024013297 |
| Model numbers | 106524INT, 106524US |
Product
Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS; 106524 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS) is a set of equipment and materials that together comprise a medical device designed to provide therapeutic benefit to those affected with advanced heart failure. In service, the LVAS assumes some or all of the workload of the left ventricle, thereby restoring the patient's systemic perfusion while palliating the underlying pathology. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus all of the features, controls, attachments, interfaces, power sources, supporting equipment, labeling, and tools required to achieve the desired therapeutic benefit. Component: No
Reason for recall
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.