Data Foundry

Medical device recalls 2012

Siemens Healthcare BNP — Class II medical device recall Z-0506-2013

Terminated recall by Siemens Healthcare Diagnostics, Inc, reported 2012-12-19, distributed nationwide. Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature

Recall numberZ-0506-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc, East Walpole, MA, United States
Recall initiated2012-11-14
Classified2012-12-11
Reported by FDA2012-12-19
Terminated2015-09-25
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BE, BR, CA, CH, CL, CN, CO, CZ, DE, DK, EE, EG, ES, FI, FR, GB, GR, HK, HR, HU, ID, IE, IN, IR, IS, IT, JP, KR, LT, LV, MX, MY, NL, NO, OM, PH, PL, PT, QA, RO, RS, SA, SE, SI, SK, TH, TR, TW, UY, VN, ZA
Reason classespotency

Product

Siemens Healthcare BNP (8-type Natriuretic Peptide) 500 Test kit 02816634 100 Test 02816138 10309045, 10309044 028 Intended Use For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur and ADVIA Centaur XP systems. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of the severity of heart failure. In patients with acute coronary syndromes (ACS), this test, in conjunction with other known risk factors, can also be used to predict survival as well as to predict the likelihood of future heart failure.

Reason for recall

Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0506-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.