Data Foundry

Medical device recalls 2012

Siemens Healthcare BR Assay for CA 27 — Class II medical device recall Z-0507-2013

Terminated recall by Siemens Healthcare Diagnostics, Inc, reported 2012-12-19, distributed nationwide. Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature

Recall numberZ-0507-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc, East Walpole, MA, United States
Recall initiated2012-11-14
Classified2012-12-11
Reported by FDA2012-12-19
Terminated2015-09-25
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BE, BR, CA, CH, CL, CN, CO, CZ, DE, DK, EE, EG, ES, FI, FR, GB, GR, HK, HR, HU, ID, IE, IN, IR, IS, IT, JP, KR, LT, LV, MX, MY, NL, NO, OM, PH, PL, PT, QA, RO, RS, SA, SE, SI, SK, TH, TR, TW, UY, VN, ZA
Reason classespotency

Product

Siemens Healthcare BR Assay for CA 27.29 250 Test Kits 03896216 50 Test Kit 02419937 Ref Kit (US only) 10340081 Intended Use The ADVIA Centaur BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur and ADVIA Centaur XP systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage Ill breast cancer.

Reason for recall

Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.