The Biomeme SARS-CoV-2 Real-Time RT-PCR Test is a molecular in vitro… — Class II medical device recall Z-0513-2021

Terminated recall by Biomeme, Inc., reported 2020-12-09, distributed nationwide. The firm has become aware of nine reports by a single customer that the software made false positive calls. Th

Recall numberZ-0513-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBiomeme, Inc., Philadelphia, PA, United States
Recall initiated2020-10-23
Classified2020-12-01
Reported by FDA2020-12-09
Terminated2021-01-25
DistributedNationwide (US)
Countries outside the USCA, DO, SG
Quantity105
Reason classespotency, software

Product

The Biomeme SARS-CoV-2 Real-Time RT-PCR Test is a molecular in vitro diagnostic test that is based on widely used nucleic acid amplification technology. The Biomeme SARS-CoV-2 Real-Time RT-PCR Test contains primers and probes and internal controls used in RT-PCR for the in vitro qualitative detection of SARS-CoV-2 RNA in upper respiratory specimens. The Biomeme SARS-CoV-2 Real-Time RT-PCR Test utilizes Biomemes M1 Sample Prep Cartridge for RNA extraction, Biomemes SARS-CoV-2 Go-Strips assay, and Biomemes portable Franklin" Real-Time qPCR Thermocycler. Franklins companion mobile app, Biomeme Go, scans tests, runs PCR experiments online or offline, and is used to quickly interpret your test results while conveniently syncing data to the Biomeme Cloud. Biomeme: 1000013 (Android Smartphone w/ Biomeme Go Mobile App) and Biomeme: 1000012 (Rugged Android Smartphone w/ Biomeme Go Mobile App) are controller for Biomeme Franklin thermocycler.

Reason for recall

The firm has become aware of nine reports by a single customer that the software made false positive calls. The investigation determined that the root cause is the higher than expected amber-to-red co-excitation in the assay.