Patient Circuit used with the Life Pulse High Frequency Ventilator — Class I medical device recall Z-0514-2013
Terminated recall by Bunnell, Inc., reported 2012-12-26, distributed nationwide. Customer complaints received indicate the heater wire insulation can melt, causing a short which creates spark
| Recall number | Z-0514-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Bunnell, Inc., Salt Lake City, UT, United States |
| Recall initiated | 2012-11-19 |
| Classified | 2012-12-14 |
| Reported by FDA | 2012-12-26 |
| Terminated | 2014-02-21 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, MY, UY |
| Quantity | 5,743 patient |
| Model numbers | 397, 902, 937 |
Product
Patient Circuit used with the Life Pulse High Frequency Ventilator. The Patient Circuit is a required disposable component of the Life Pulse High Frequency Ventilator. Model #(s): Individual Patient Circuit - catalog # 902 Patient Circuit Kit - catalog # 937: contains 2 circuits and 2 x 2.5 & 2 x 3.5 LifePort Adapters. Labeling states manufactured by: Bunnell Incorporated 436 Lawndale Drive, Salt Lake City, Utah 84115. Used for ventilating critically ill infants with pulmonary interstitial emphysema and infants with respiratory distress syndrome complicated by pulmonary air leaks, who are, in the opinion of their physicians, failing on conventional ventilation. The Patient Circuit is to provide a conduit for the humidification, warming, and temperature monitoring of the pressurized gas. The Patient Circuit is indicated for seven day single use.
Reason for recall
Customer complaints received indicate the heater wire insulation can melt, causing a short which creates sparking and smoke, in the circuit close to the humidifier cartridge. Bunnell is recalling all lots of circuits distributed between March 12, 2012 and November 30, 2012.