Data Foundry

Medical device recalls 2020

Invacare PerfectO2 V Oxygen Concentrator- intended for patients with… — Class II medical device recall Z-0514-2021

Terminated recall by Invacare Corporation, reported 2020-12-09, distributed nationwide. Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve

Recall numberZ-0514-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmInvacare Corporation, Elyria, OH, United States
Recall initiated2020-10-30
Classified2020-12-01
Reported by FDA2020-12-09
Terminated2024-12-31
DistributedNationwide (US)
Countries outside the USAU, CA, CN, DE, NZ
Quantity384,767 units
UPC / GTIN / UDI-DI00841447100249
Serial numbers00841447100249, 14HF039890, 14IF004028, 17HF030338-18IF018522, 17JF002563-19AF008559, 18BF007341-18IF024451, 18EF003466-18EF003544, 18JF007415, 18JF028633, IRC5PO2VAW, IRC5PO2VAWL2, IRC5PO2VAWL3, IRC5PO2VAWL5, IRC5PO2VC
Model numbersIRC5PO2V

Product

Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen Model Numbers: IRC5PO2V (US), IRC5PO2VC (Canada), All World: IRC5PO2VAW, IRC5PO2VAWL2, IRC5PO2VAWL3, IRC5PO2VAWL4, IRC5PO2VAWL5

Reason for recall

Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0514-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.