Data Foundry

Medical device recalls 2018

ADVIA Centaur Calibrator 30 — Class II medical device recall Z-0515-2018

Terminated recall by Siemens Healthcare Diagnostics, Inc, reported 2018-02-14. Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the

Recall numberZ-0515-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc, East Walpole, MA, United States
Recall initiated2017-06-07
Classified2018-02-04
Reported by FDA2018-02-14
Terminated2019-05-08
Quantity5,992 units
Lot numbers00387A21, 00388A21, 15448A21, 16673A21, 20640A21, 36243A21, 51024A21, 51211A21
Expiration dates2018-03-08

Product

ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.

Reason for recall

Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021) as compared to Calibrator 30 kit lots ending in 20 (C3020).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0515-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.