Varian On-Board Imager Advanced Imaging System — Class II medical device recall Z-0526-2015

Terminated recall by Varian Medical Systems, Inc., reported 2014-12-24. Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were ship

Recall numberZ-0526-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmVarian Medical Systems, Inc., Palo Alto, CA, United States
Recall initiated2014-04-30
Classified2014-12-16
Reported by FDA2014-12-24
Terminated2015-03-16
Quantity32 units
Model numbersP1003290001

Product

Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.

Reason for recall

Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.