Optional MDRC-1119 Front Cover Attachment Kit for review display… — Class II medical device recall Z-0527-2019
Terminated recall by Barco N.V., reported 2018-12-12, distributed nationwide. If metal strips are removed and re-applied to another display, there is potential for the metal strips to be d
| Recall number | Z-0527-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Barco N.V., Kortrijk, N/A, Belgium |
| Recall initiated | 2017-03-30 |
| Classified | 2018-11-30 |
| Reported by FDA | 2018-12-12 |
| Terminated | 2020-05-04 |
| Distributed | Nationwide (US) |
| Countries outside the US | BE, BR, CA, CZ, DE, DK, ES, FI, FR, IT, JP, KR, PL, SA, SE, TR, TW, ZA |
| Quantity | 2,087 units |
| Model numbers | C9826096, C9826097, C9826108, C9826109 |
Product
Optional MDRC-1119 Front Cover Attachment Kit for review display monitors. Product Usage: The product is an optional kit, intended to be attached in front of the LCD panel of a review display (MDRC-1119).
Reason for recall
If metal strips are removed and re-applied to another display, there is potential for the metal strips to be deformed resulting in a smaller contact surface with the front cover magnet strips. There is a risk that the protective front cover could come loose and may fall down.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0527-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.