Varian brand ARIA Radiation Oncology — Class II medical device recall Z-0531-2013

Terminated recall by Varian Medical Systems, Inc. Oncology Systems, reported 2012-12-26. Varian Medical Systems has identified an anomaly with the ARIA Oncology Information System Radiation Oncology

Recall numberZ-0531-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmVarian Medical Systems, Inc. Oncology Systems, Palo Alto, CA, United States
Recall initiated2012-11-05
Classified2012-12-17
Reported by FDA2012-12-26
Terminated2014-03-03
Quantity264 units

Product

Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application, Model Number: HIT, Affected Versions: ARIA version 11.0 below 11.0.55, Build 11.0.28 and 11.0.34, Reference/FSCA Identifier: CP-09798; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA

Reason for recall

Varian Medical Systems has identified an anomaly with the ARIA Oncology Information System Radiation Oncology [ARIA RO] DICOM import/export functionality where a 360 degree arc field may convert to static field during DICOM export or import.