Accu2i pMTA Applicator — Class II medical device recall Z-0531-2018
Terminated recall by Angiodynamics, Inc., reported 2018-02-14, distributed nationwide. The firm is recalling product that has the potential to not function during use. The product affected by this
| Recall number | Z-0531-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Angiodynamics, Inc., Queensbury, NY, United States |
| Recall initiated | 2017-06-05 |
| Classified | 2018-02-05 |
| Reported by FDA | 2018-02-14 |
| Terminated | 2018-09-11 |
| Distributed | Nationwide (US) |
| Quantity | 14,353 units |
| Lot numbers | 15330420, 15370436, 15390448, 15430457, 15450464, 15480474, 15490479, 15520485, 16030491, 16140518, 16160527, 16180533, 16300570, 16320576, 16350584, 16440615, 16460622, 16480628, 16500640, 16510643, 17010648, 17030656 |
| Model numbers | 806-10, 900-602-US, H787900602US0 |
Product
Accu2i pMTA Applicator, 19 cm INTERMEDIATE, REF/Catalog # 900-602-US, Model # (UPN) H787900602US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.
Reason for recall
The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0531-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.