Data Foundry

Medical device recalls 2024

VITROS Chemistry Products CRP Slides- In vitro diagnostic… — Class II medical device recall Z-0535-2025

Ongoing recall by Ortho-Clinical Diagnostics, Inc., reported 2024-12-04, distributed nationwide. May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than

Recall numberZ-0535-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Inc., Rochester, NY, United States
Recall initiated2024-10-11
Classified2024-11-25
Reported by FDA2024-12-04
DistributedNationwide (US)
Countries outside the USAU, BE, BM, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG
Quantity7,717 units
UPC / GTIN / UDI-DI10758750000296
Model numbers1926740
Expiration dates2024-10-01, 2025-01-01

Product

VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740

Reason for recall

May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0535-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.