REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer… — Class II medical device recall Z-0542-2026

Ongoing recall by Cellavision AB, reported 2025-11-26, distributed nationwide. Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting

Recall numberZ-0542-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCellavision AB, Lund, Sweden
Recall initiated2025-10-08
Classified2025-11-18
Reported by FDA2025-11-26
DistributedNationwide (US)
Quantity688 units
Serial numbers60001-63219, 63220-70093, CN60001-CN63219, CN63220-CN70093
Model numbersBS737162, CC286297

Product

REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest or Jadk FM-5 barcode reader component.

Reason for recall

Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.