Data Foundry

Medical device recalls 2023

Philips Allura systems with monoplane fixed ceiling mounts: 722001… — Class II medical device recall Z-0558-2024

Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2023-12-27, distributed nationwide. Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision

Recall numberZ-0558-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, N/A, Netherlands
Recall initiated2023-11-03
Classified2023-12-15
Reported by FDA2023-12-27
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CW, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, KY, LB, LK, LT, LV, MA, MX, MY, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, UY, VE, VN, ZA
Quantity9,991 units
Reason classessterility
UPC / GTIN / UDI-DI00884838054189, 00884838054202, 00884838054240, 00884838059030, 00884838059054, 00884838059085, 00884838059115

Product

Philips Allura systems with monoplane fixed ceiling mounts: 722001 Allura Xper FD10C 722003 Allura Xper FD10 722006 Allura Xper FD20 722010 Allura Xper FD10 722012 Allura Xper FD20 722015 Allura Xper FD20 OR Table 722016 Integris H5000 C / Allura 9C 722018 Integris Allura 9 722022 Allura Xper FD10 OR Table 722023 Allura Xper FD20 OR Table 722026 Allura Xper FD10 722028 Allura Xper FD20 722033 Allura Xper FD10 OR Table 722035 Allura Xper FD20 OR Table 722043 Integris Allura 15 & 12 (monoplane) 72246 Poly G - OMCP - VISUB - CCD (H5000)

Reason for recall

Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0558-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.