Data Foundry

Medical device recalls 2016

MicroScan Pos Breakpoint Combo Panel Type 23 — Class II medical device recall Z-0562-2016

Terminated recall by Beckman Coulter Inc., reported 2016-01-06, distributed nationwide. Beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) wi

Recall numberZ-0562-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2015-12-01
Classified2015-12-30
Reported by FDA2016-01-06
Terminated2016-07-29
DistributedNationwide (US)
Countries outside the USAT, BE, BR, CA, DE, ES, GB, IT, JP, KW, MX, NL, NO, PA, PL, PT, RO, TH, TW
Pathogensstaphylococcus
Model numbersB1017-206

Product

MicroScan Pos Breakpoint Combo Panel Type 23, Part No. B1017-206 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.

Reason for recall

Beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) with Quality Control (QC) American Type Culture Collection (ATCC) organism Staphylococcus aureus ATCC 29213 in a portion of affected lots of MicroScan Pos Combo and MicroScan Pos Breakpoint Combo panels. The expected result is positive.