Data Foundry

Medical device recalls 2024

(1) Nurse Kit, SKU #EDUC1025, Case UPC 40195327196647, Non-sterile,… — Class I medical device recall Z-0563-2024

Ongoing recall by MEDLINE INDUSTRIES, LP - Northfield, reported 2024-01-03, distributed nationwide. The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use a

Recall numberZ-0563-2024
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMEDLINE INDUSTRIES, LP - Northfield, Northfield, IL, United States
Recall initiated2023-10-16
Classified2023-12-22
Reported by FDA2024-01-03
DistributedNationwide (US)
Countries outside the USAE, CA, PA
Quantity550 kits
Reason classesdevice malfunction
UPC / GTIN / UDI-DI40195327153695, 40195327196647
Lot numbers22HDA420, 22HDC201, 22HDC340, 22JDB315

Product

(1) Nurse Kit, SKU #EDUC1025, Case UPC 40195327196647, Non-sterile, For Training Purposes Only, Not For Human Use.; and (2) Nursing Tote Kit, SKU #EDUC5056A, Case UPC 40195327153695, Non-sterile, For Training Purposes Only, Not For Human Use.

Reason for recall

The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0563-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.