Data Foundry

Medical device recalls 2014

Instrument Manager driver : Driver sysu100i that connects to the… — Class II medical device recall Z-0571-2014

Terminated recall by Data Innovations, Inc., reported 2014-01-01, distributed nationwide. When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001

Recall numberZ-0571-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmData Innovations, Inc., South Burlington, VT, United States
Recall initiated2013-11-06
Classified2013-12-24
Reported by FDA2014-01-01
Terminated2014-12-05
DistributedNationwide (US)
Countries outside the USCA
Quantity24 systems
Reason classessoftware

Product

Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004

Reason for recall

When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error