Data Foundry

Medical device recalls 2018

Customer Epidural Anesthesia Tray — Class II medical device recall Z-0574-2018

Terminated recall by B. Braun Medical, Inc., reported 2018-02-21, distributed nationwide. B. Braun Medical Inc. (BBMI) initiated this recall of various BBMI convenience kits containing the 22GA 1-1/2i

Recall numberZ-0574-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmB. Braun Medical, Inc., Allentown, PA, United States
Recall initiated2017-06-21
Classified2018-02-09
Reported by FDA2018-02-21
Terminated2018-04-03
DistributedNationwide (US)
Quantity120 units
Reason classesforeign material
Model numbers555234

Product

Customer Epidural Anesthesia Tray. Product Catalog Number: 555234. Local anesthesia of the skin prior to insertion of the regional anesthesia needle.

Reason for recall

B. Braun Medical Inc. (BBMI) initiated this recall of various BBMI convenience kits containing the 22GA 1-1/2in SafetyGlide Needle due to the presence of loose polypropylene foreign matter was above release specification. This foreign matter has been identified as a product hub material which has been tested for biocompatibility per ISO 10993 during the product development process.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0574-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.