Data Foundry

Medical device recalls 2018

EMS products from Graham Medical — Class II medical device recall Z-0578-2018

Terminated recall by LITTLE RAPIDS CORPORATION, reported 2018-02-21, distributed in CA, CT, FL, MO, MS, SD, TN, TX…. The secondary supplier provided material was within specification but did not meet our design needs compared t

Recall numberZ-0578-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmLITTLE RAPIDS CORPORATION, Green Bay, WI, United States
Recall initiated2017-12-27
Classified2018-02-09
Reported by FDA2018-02-21
Terminated2018-07-30
DistributedCA, CT, FL, MO, MS, SD, TN, TX, WI
Lot numbers17871, 17971, 18071, 18971

Product

EMS products from Graham Medical, SnugFit(R) Expandable Mattress Cover, Code 72930, UPC 1 07 47036 72930 3

Reason for recall

The secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient transfer, not lifting. However, we recognize that the product may be utilized outside of its intended design scope, and lifting may occur. Lifting could lead tearing of the equipment cover and the potential for patient falling through.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0578-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.