EMS products from Graham Medical — Class II medical device recall Z-0578-2018
Terminated recall by LITTLE RAPIDS CORPORATION, reported 2018-02-21, distributed in CA, CT, FL, MO, MS, SD, TN, TX…. The secondary supplier provided material was within specification but did not meet our design needs compared t
| Recall number | Z-0578-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | LITTLE RAPIDS CORPORATION, Green Bay, WI, United States |
| Recall initiated | 2017-12-27 |
| Classified | 2018-02-09 |
| Reported by FDA | 2018-02-21 |
| Terminated | 2018-07-30 |
| Distributed | CA, CT, FL, MO, MS, SD, TN, TX, WI |
| Lot numbers | 17871, 17971, 18071, 18971 |
Product
EMS products from Graham Medical, SnugFit(R) Expandable Mattress Cover, Code 72930, UPC 1 07 47036 72930 3
Reason for recall
The secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient transfer, not lifting. However, we recognize that the product may be utilized outside of its intended design scope, and lifting may occur. Lifting could lead tearing of the equipment cover and the potential for patient falling through.