Data Foundry

Medical device recalls 2024

Ziehm Vision RFD — Class II medical device recall Z-0578-2025

Ongoing recall by Orthoscan, Inc., reported 2024-12-11, distributed in PR. Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed device

Recall numberZ-0578-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedFDA mandated
Recalling firmOrthoscan, Inc., Scottsdale, AZ, United States
Recall initiated2024-08-08
Classified2024-12-02
Reported by FDA2024-12-11
DistributedPR
Quantity30
Reason classesdevice malfunction
Serial numbers20048, 20240, 21350, 23249, 23250, 23251, 23252, 23253, 23324, 23375, 23376, 23377, 23378, 23379, 23380, 23381, 23382, 23412, 23452, 23460, 23461, 23462, 23463, 23506, 23536 and 5 more

Product

Ziehm Vision RFD. Interventional fluoroscopic x-ray system

Reason for recall

Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.