Data Foundry

Medical device recalls 2019

Evolution Loading Car and Transfer Carriage- Accessory to the… — Class II medical device recall Z-0586-2020

Terminated recall by Steris Corporation, reported 2019-12-11, distributed nationwide. The locking pin of the front wheel asse mbly of the Evolution Transfer Carriages may not remain full y engaged

Recall numberZ-0586-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSteris Corporation, Mentor, OH, United States
Recall initiated2019-10-22
Classified2019-11-29
Reported by FDA2019-12-11
Terminated2020-06-09
DistributedNationwide (US)
Countries outside the USBR, CA, MX, PE, SA
Quantity237 units
Serial numbers30021901, 30031931, 30031932, 30031933, 30031934, 30031935, 30091907, 30091908, 30101924, 30101925, 30101926, 30101927, 30101928, 30101929, 30101930, 30121904, 30121905, 30121906, 30151916, 30151917, 30151918, 30151919, 30151920, 30161906, 30161907 and 212 more

Product

Evolution Loading Car and Transfer Carriage- Accessory to the Sterilizer, Steam

Reason for recall

The locking pin of the front wheel asse mbly of the Evolution Transfer Carriages may not remain full y engaged with the transfer carriage and result in the transfer carriage and loading car unexpectedly falling forward

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0586-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.