MOVES System Lithium Polymer Batteries The MOVES System is a portable… — Class I medical device recall Z-0587-2016
Terminated recall by Thornhill Research Inc, reported 2016-01-20, distributed in VA. Batteries found unresponsive after storage, not recognized by main device, and are perpetually enabled.
| Recall number | Z-0587-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Thornhill Research Inc, Toronto, N/A, Canada |
| Recall initiated | 2015-11-19 |
| Classified | 2016-01-11 |
| Reported by FDA | 2016-01-20 |
| Terminated | 2019-04-12 |
| Distributed | VA |
| Countries outside the US | SG |
| Quantity | 178 units |
Product
MOVES System Lithium Polymer Batteries The MOVES System is a portable computer controlled electrically powered transport emergency ventilator intended to provide continuous or intermittent ventilatory support for the care of adults who require ventilation.
Reason for recall
Batteries found unresponsive after storage, not recognized by main device, and are perpetually enabled.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0587-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.