Quantum Intersegmental Table — Class II medical device recall Z-0588-2026

Ongoing recall by Pivotal Health Solutions, Inc., reported 2025-12-03, distributed nationwide. The heat function may become damaged over time due to patients pressing on the areas with heat function compon

Recall numberZ-0588-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPivotal Health Solutions, Inc., Watertown, SD, United States
Recall initiated2025-10-09
Classified2025-11-24
Reported by FDA2025-12-03
DistributedNationwide (US)
Countries outside the USCA, JP
Quantity464 units
Reason classesdevice malfunction
Model numbersAMQ400H

Product

Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table

Reason for recall

The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.