Data Foundry

Medical device recalls 2016

Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz… — Class II medical device recall Z-0589-2016

Terminated recall by Maquet Cardiovascular Us Sales, Llc, reported 2016-01-13, distributed nationwide. Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can caus

Recall numberZ-0589-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMaquet Cardiovascular Us Sales, Llc, Wayne, NJ, United States
Recall initiated2015-11-13
Classified2016-01-05
Reported by FDA2016-01-13
Terminated2017-04-07
DistributedNationwide (US)
Quantity100 units
Reason classesdevice malfunction
Model numbers704629/70103.4371HCU, 704631/70103.4642HCU, 705501/70103.4653

Product

Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

Reason for recall

Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the compressor is switched on (prior to patient use); however, it can occur during patient use. This can shut down the compressor or the entire HCU 30 Type 2 unit.