Data Foundry

Medical device recalls 2018

CORAIL COXA VARA HIGH OFFSET STEM Collared Size 9 — Class II medical device recall Z-0593-2018

Terminated recall by DePuy Orthopaedics, Inc., reported 2018-02-21. Incorrect device in the package. A package that was labeled as the CORAIL COXA VARA HIGH OFFSET STEM Collared

Recall numberZ-0593-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDePuy Orthopaedics, Inc., Warsaw, IN, United States
Recall initiated2017-07-12
Classified2018-02-12
Reported by FDA2018-02-21
Terminated2019-05-20
Quantity21 units
Reason classeslabeling
Lot numbers5291990
Model numbers3L93709

Product

CORAIL COXA VARA HIGH OFFSET STEM Collared Size 9

Reason for recall

Incorrect device in the package. A package that was labeled as the CORAIL COXA VARA HIGH OFFSET STEM Collared Size 9 (Part No. 3L93709 / Lot No. 5291990), contained the CORAIL HIGH OFFSET STEM Collarless SIZE 14 (Part No. L20314 / Lot No. 5292130).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0593-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.