AIA-2000 Analyzer — Class II medical device recall Z-0599-2019
Terminated recall by Tosoh Bioscience Inc, reported 2018-12-19, distributed nationwide. Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could resu
| Recall number | Z-0599-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Tosoh Bioscience Inc, Grove City, OH, United States |
| Recall initiated | 2018-09-28 |
| Classified | 2018-12-13 |
| Reported by FDA | 2018-12-19 |
| Terminated | 2020-09-03 |
| Distributed | Nationwide (US) |
| Countries outside the US | CL, CO, CR, EC, GT, HN, KY, PA, PE, UY, VE |
| Quantity | 95 |
Product
AIA-2000 Analyzer, Product code 022100, 022100R, 022101, 022101R The AIA-2000 analyzer and AIA-900 analyzer are capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step immunoenzymetric immunoassay.
Reason for recall
Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue.