epoc NXS Host Blood Analysis System — Class II medical device recall Z-0601-2024

Ongoing recall by Siemens Healthcare Diagnostics Inc, reported 2023-12-27, distributed nationwide. Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions,

Recall numberZ-0601-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics Inc, Norwood, MA, United States
Recall initiated2023-10-19
Classified2023-12-20
Reported by FDA2023-12-27
DistributedNationwide (US)
Quantity10,183 devices
Reason classessoftware
UPC / GTIN / UDI-DI00630414605678, 00630414605760, 00630414605814, 00630414605821, 00630414605838, 00630414606095, 00630414612447, 00630414631028

Product

epoc NXS Host Blood Analysis System, Siemens Material Numbers 11413497 (EU), 11413475 (US), 11413506 (CA), 11413517 (MX), 11413518 (ROW), 11413498 (JP), 11413583 (CN, KR), 11413879 (IN) with software versions from 4.9.4 to the current version 4.11.11

Reason for recall

Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions, a software problem is causing erroneous results to be printed. When the results are run on the epoc NXS Host and Delete Blood Tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. The worst possible outcome may result in a failure to diagnose hyperkalemia, due to an erroneous depressed potassium level or inappropriate treatment to hyperglycemia due to erroneous glucose reading.