IntelliFill i — Class II medical device recall Z-0607-2014
Terminated recall by Baxter Healthcare Corp., reported 2014-01-08, distributed nationwide. Following a retrospective review for Baxter Daytona Complaint CR-954, it was identified that the IntelliFill i
| Recall number | Z-0607-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Baxter Healthcare Corp., Deerfield, IL, United States |
| Recall initiated | 2013-11-08 |
| Classified | 2013-12-29 |
| Reported by FDA | 2014-01-08 |
| Terminated | 2015-03-23 |
| Distributed | Nationwide (US) |
| Quantity | 28 devices |
| Lot numbers | 123, 125R, 127, 129, 145, 146R, 147R, 148, 215R, 500, 501, 502, 503, 504, 505, 506, 508, 510, 511R, 512, 514, 515, 516R, 517, 518, 519, 520 |
Product
IntelliFill i.v. Pharmacy Compounding Device (PCD). Intended to fill syringes from 0.5mL up to 11.5mL using mixing instructions from its formulary to fill the orders.
Reason for recall
Following a retrospective review for Baxter Daytona Complaint CR-954, it was identified that the IntelliFill i.v. Pharmacy Compounding Device (PCD) (product code: IFIV) has the potential to continue to run and fill syringes with the HEPA filtration disabled.