Data Foundry

Medical device recalls 2013

Access Immunoassay System — Class II medical device recall Z-0612-2013

Terminated recall by Beckman Coulter Inc., reported 2013-01-02, distributed nationwide. It has been determined that, in some cases, pulley drive components have broken parts or have become loose, mi

Recall numberZ-0612-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2012-11-01
Classified2012-12-22
Reported by FDA2013-01-02
Terminated2013-12-23
DistributedNationwide (US)
Countries outside the USCA
Quantity6,858 units

Product

Access Immunoassay System, Access 2 Immunoassay System, Synchron LXi 725 Clinical System, UniCel DxC 600i Synchron Access Clinical System Part Numbers: Access  81600; Access 2  81600N, 386220, A69186, A25656, A12900; Access 2 (LXi)  A15642; Access 2 Section, DxC 600i  A25640. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The Synchron LXi and the UniCel DxC 600i are Access systems integrated with a chemistry analyzer.

Reason for recall

It has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at an incorrect speed, resulting in inadequate washing or mixing of the contents of the reaction vessel.