Philips HeartStart MRx Monitor/Defibrillator Models: M3536A — Class II medical device recall Z-0613-2014

Terminated recall by Philips Medical Systems, Inc., reported 2014-01-08, distributed nationwide. Device Operating on Battery Power May Shutdown without Warning if exposed to elevated levels of electromagneti

Recall numberZ-0613-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Medical Systems, Inc., Andover, MA, United States
Recall initiated2013-04-10
Classified2013-12-31
Reported by FDA2014-01-08
Terminated2017-04-13
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, AW, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HR, ID, IE, IL, IN, IQ, IS, IT, JO, JP, KE, KR, KW, LB, LK, LT, LU, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SV, TH, TN, TR, TT, TW, UA, VE, VN, ZA
Quantity64,972
Model numbersM3536A, M3536J, M3536M, M3536M2, M3536M4, M3536M5, M3536M6, M3536MC

Product

Philips HeartStart MRx Monitor/Defibrillator Models: M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician

Reason for recall

Device Operating on Battery Power May Shutdown without Warning if exposed to elevated levels of electromagnetic interference from other radio frequency (RF) energy sources and