Alinity ci-series System Control Module — Class II medical device recall Z-0613-2024
Ongoing recall by Abbott Laboratories, reported 2024-01-03, distributed in AK, AL, AR, AZ, CA, CO, CT, DE…. There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower
| Recall number | Z-0613-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abbott Laboratories, Irving, TX, United States |
| Recall initiated | 2023-05-30 |
| Classified | 2023-12-22 |
| Reported by FDA | 2024-01-03 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BG, BH, BM, BR, BS, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IT, JM, JO, JP, KE, KR, KW, KY, LB, LT, LV, MA, MX, MY, NI, NL, NO, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VN, ZA |
| Quantity | 8,424 |
| Reason classes | labeling, potency, software |
| UPC / GTIN / UDI-DI | 00380740137410 |
| Model numbers | 03R70-01 |
Product
Alinity ci-series System Control Module, REF: 03R70-01
Reason for recall
There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3) Customer requests mechanism to prevent user error when creating new lots of value assigned Calibrators; 4) Error in processing tests after CCCintController encountered an error with the assay information downloaded from SCC; 5) The assay editor does not correctly check sample dilution total volume limits; 6) User-Applied Labels not adhering to Reagent Bottles; 7) ICT Module did not expire after warranty was exceeded. These performance issues could lead to erroneous results for multiple analytes.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0613-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.