Data Foundry

Medical device recalls 2013

Medical Surgical Bed — Class II medical device recall Z-0629-2013

Terminated recall by Stryker Medical Division of Stryker Corporation, reported 2013-01-09, distributed nationwide. Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. may have

Recall numberZ-0629-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Medical Division of Stryker Corporation, Portage, MI, United States
Recall initiated2012-12-18
Classified2013-01-03
Reported by FDA2013-01-09
Terminated2013-08-30
DistributedNationwide (US)
Countries outside the USCA
Quantity1,004
Serial numbers120415087, 120515468
Model numbers3005, 3005S3EX

Product

Medical Surgical Bed, Model 3005S3EX Medical Surgical Bed The S3" MedSurg Bed, Model 3005 is intended to support and transport patients within the Med/Surg and Critical Care hospital environments. The bed is typically used in pre-op, post-op and recovery areas of hospital facilities. The intended user for this product is both Health Care Providers (HCPs: nurses, nurses aides, and medical doctors) and human patients. Lockout features may limit patient accessible controls.

Reason for recall

Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0629-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.