Data Foundry

Medical device recalls 2022

Arrow-Trerotola PTD Set — Class I medical device recall Z-0629-2022

Terminated recall by TELEFLEX MEDICAL INC, reported 2022-02-23, distributed nationwide. The firm received reports indicating PTD tip separation during use.

Recall numberZ-0629-2022
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTELEFLEX MEDICAL INC, Morrisville, NC, United States
Recall initiated2022-01-11
Classified2022-02-15
Reported by FDA2022-02-23
Terminated2024-09-11
DistributedNationwide (US)
Quantity14,583 units
UPC / GTIN / UDI-DI00801902013156, 10801902013153, 10801902194074, 20801902194071
Model numbersPT-65509

Product

Arrow-Trerotola PTD Set (Percutaneous Thrombolytic Device), REF PT-65509 (finished good kit) The Arrow-Trerotola Percutaneous Thrombolytic Device (PTD) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venous (AV) fistulae and synthetic dialysis grafts.

Reason for recall

The firm received reports indicating PTD tip separation during use.