Data Foundry

Medical device recalls 2013

Siemens CLINITEK Atlas Positive and Negative Control Strips for Urine… — Class III medical device recall Z-0636-2013

Terminated recall by Siemens Healthcare Diagnostics Inc, reported 2013-01-16, distributed nationwide. When these lots of control material are reconstituted, the positive control solution leukocyte result may beco

Recall numberZ-0636-2013
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics Inc, Norwood, MA, United States
Recall initiated2012-11-21
Classified2013-01-04
Reported by FDA2013-01-16
Terminated2014-11-06
DistributedNationwide (US)
Countries outside the USAW, BR, BS, CA, DK, HK, IN, MX, NZ, SG, TH
Quantity6,466
Lot numbersP0097091C, P0100101B, P0103042A, P0103042H, P0106062A, P9093061A, P9094081A, P9101022A, P9101022B, P9105052A, S0097091C, S0100101B, S0103042A, S0103042H, S0106062A, S9093061A, S9094081A, S9101022A, S9101022B, S9105052A
Model numbers5019

Product

Siemens CLINITEK Atlas Positive and Negative Control Strips for Urine Chemistry Catalog Number: 5019 When used with the CLINITEK ATLAS¿ Automated urine Chemistry Analyzer and CLINITEK ATLAS¿ 10 reagent Pak, CLINITEK ATLAS¿ Positive and Negative Control Strips provide defined results for color, clarity, glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogen, nitrite, and leukocytes. Each CLINITEK ATLAS Positive Control Strip is a firm plastic strip to which are affixed seven separate analyte areas.

Reason for recall

When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).