Data Foundry

Medical device recalls 2014

Sandel Healthcare Safety Devices Weighted Safety Scalpel #3… — Class II medical device recall Z-0639-2014

Terminated recall by Ansell Sandel Medical Industries LLC, reported 2014-01-15, distributed nationwide. Select lots of SANDEL Weighted Safety Scalpels are packaged with the scalpel blade exposed beyond the safety s

Recall numberZ-0639-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAnsell Sandel Medical Industries LLC, Chatsworth, CA, United States
Recall initiated2013-11-12
Classified2014-01-07
Reported by FDA2014-01-15
Terminated2016-04-06
DistributedNationwide (US)
Countries outside the USCA
Reason classessterility
Model numbers1203, 1553, 2215-NLN

Product

Sandel Healthcare Safety Devices Weighted Safety Scalpel #3 Disposable Safety Handle with #15 Blade Manufactured for: Ansell Sandel Medical Solutions LLC 19736 Dearborn St. Chatsworth, CA 91311 USA Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Made in China Packaged in Mexico Ansell A manual surgical scalpel blade for making incisions in various general surgical procedures.

Reason for recall

Select lots of SANDEL Weighted Safety Scalpels are packaged with the scalpel blade exposed beyond the safety shield which could compromise the integrity of the package and the sterility of the scalpel leading to injury.