VITROS 3600 Chemistry System — Class II medical device recall Z-0640-2016

Terminated recall by Ortho-Clinical Diagnostics, reported 2016-01-20, distributed nationwide. Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatur

Recall numberZ-0640-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2015-11-03
Classified2016-01-13
Reported by FDA2016-01-20
Terminated2018-05-30
DistributedNationwide (US)
Countries outside the USAR, AU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IL, IN, IT, JP, KE, MX, PA, PH, PL, RO, RU, SG, TR, TW, VE
Reason classessoftware
UPC / GTIN / UDI-DI10758750009930
Serial numbers36000879, J36000012
Model numbers6802783

Product

VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifier Number 10758750009930, Software Product Code 6802866, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.

Reason for recall

Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.