VITROS 3600 Chemistry System — Class II medical device recall Z-0640-2016
Terminated recall by Ortho-Clinical Diagnostics, reported 2016-01-20, distributed nationwide. Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatur
| Recall number | Z-0640-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Rochester, NY, United States |
| Recall initiated | 2015-11-03 |
| Classified | 2016-01-13 |
| Reported by FDA | 2016-01-20 |
| Terminated | 2018-05-30 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IL, IN, IT, JP, KE, MX, PA, PH, PL, RO, RU, SG, TR, TW, VE |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 10758750009930 |
| Serial numbers | 36000879, J36000012 |
| Model numbers | 6802783 |
Product
VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifier Number 10758750009930, Software Product Code 6802866, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.
Reason for recall
Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0640-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.