C7000 — Class II medical device recall Z-0644-2020

Terminated recall by Integra LifeSciences Corp., reported 2019-12-18, distributed in AL, AR, AZ, CA, CO, CT, FL, GA…. The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a

Recall numberZ-0644-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmIntegra LifeSciences Corp., Plainsboro, NJ, United States
Recall initiated2019-10-21
Classified2019-12-08
Reported by FDA2019-12-18
Terminated2020-05-07
DistributedAL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, UT, VA, WA, WI, WV

Product

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Reason for recall

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.