The Vantage Digital Panoramic System delivers high-value features… — Class II medical device recall Z-0651-2014
Terminated recall by Midmark Corp dba Progeny Inc, reported 2014-01-15, distributed nationwide. An error was identified in the Real Time Controller (RTC) firmware versions 3.0, 3.1, 3.2, and 3.3 used in Van
| Recall number | Z-0651-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Midmark Corp dba Progeny Inc, Lincolnshire, IL, United States |
| Recall initiated | 2013-09-20 |
| Classified | 2014-01-09 |
| Reported by FDA | 2014-01-15 |
| Terminated | 2017-12-27 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, RU |
| Quantity | 58 units |
| Reason classes | software |
| Serial numbers | Z08018, ZB51001, ZB51002, ZB51006, ZB51007, ZE09006, ZE09012, ZE09019, ZK58019, ZM81016, ZM81017, ZM81022, ZM81023, ZM81025, ZM81026, ZM81027, ZM81028, ZM81030, ZN44018, ZN44020, ZP01043, ZS54011, ZS54012, ZS59001, ZS59007 and 33 more |
| Model numbers | V5000, V5050, V5100, V5150, V5150C |
Product
The Vantage Digital Panoramic System delivers high-value features standard on every system: - Five pre-programmed exam settings, including bitewing, minimize exam set-up time and allow routine diagnostic exams to be performed extraorally for improved clinical efficiency and patient experience. - Progeny Clarity Enhanced Filter Set provides superior diagnostic imaging quality - Exclusive Focal Trough Alignment reminder assists in consistent patient positioning - Progeny Imaging software included for download on an unlimited number of operatory workstations with no additional costs - Exclusive VantageTrust remote support and training service offered at no charge for the first 60 days of use - Ceph upgradeable for expanded extraoral exam capabilities - The Progeny Vantage Panoramic System s proudly designed and manufactured in the USA; 2013 Midmark Corporation; 675 Heathrow Drive Lincolnshire, IL 60069; 847-415-9800 Toll-free 888-924-3800
Reason for recall
An error was identified in the Real Time Controller (RTC) firmware versions 3.0, 3.1, 3.2, and 3.3 used in Vantage Panoramic X-Ray systems. The error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. failure to stop").
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0651-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.