Data Foundry

Medical device recalls 2014

***REF BLACKMAX-N***Swivel/Angle Motor Assembly***Rx Only***The… — Class II medical device recall Z-0654-2014

Terminated recall by The Anspach Effort, Inc., reported 2014-01-15, distributed nationwide. The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System,

Recall numberZ-0654-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmThe Anspach Effort, Inc., Palm Beach Gardens, FL, United States
Recall initiated2013-04-25
Classified2014-01-09
Reported by FDA2014-01-15
Terminated2015-07-20
DistributedNationwide (US)
Quantity2,785
Serial numbers31496602

Product

***REF BLACKMAX-N***Swivel/Angle Motor Assembly***Rx Only***The Anspach Effort, Inc. Intended for cutting and shaping bone.

Reason for recall

The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the hand piece's outer hose can rupture without warning if the air flow is blocked during use..

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0654-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.